Home >  Tests A–Z > Parameter
Laboratory parameter index

Vitamin A

material / volume

material / volume

serum

volume

volume

1 ml

comments

comments

serum should be obtained asap and kept dark

referece Values

referece Values

0-7 years 0,2-0,43 mg/l
7-13 years 0,26-0,49 mg/l
13-20 years 0,26-0,72 mg/l
>20 years 0,3-0,8 mg/l

Method

Method

HPLC

stability

stability

-20°C 2 weeks

TAT

TAT

10 working days

 

Methode mit ° = Fremdversand